Usage ================================== documentation of the (advanced) usage of RadIIS: SCPI commands: .. function:: SYStem:BIAS Set the SiPM bias, given in [mV]. Higher bias means a higher signal amplification. If a very low bias is set it might be impossible to digitize the signals. It is not possible to choose a bias that damages the system if the enclosure is intact. *Maximum = 29950 mV* *Minimum = 24304 mV* *Default = ?* Example: SYS:BIAS 26000 (Sets the bias to 26.00V). .. function:: SYStem:BIAS? *Returns:* SiPM bias in mV [INT] .. function:: SYStem:ATC Activates or deactivates the **A**utomatic **T**emperature **C**ompensation. If enabled the bias of the SiPM is automatically adjusted to achieve a constant gain during temperature variations. It is very recommended to acctivate the ATC during measurements. *Default = ?* Example: SYS:ATC ON (Activates the ATC). .. function:: SYStem:ATC? *Returns:* State of the ATC [BOOL] .. function:: SYStem:TEMPerature? *Returns:* SiPM temperature in mC [INT] .. function:: SYStem:COMPerator:THReshold Set the comperator threshold for the external and internal channel. The number is given in DAC counts. Higher numbers mean a higher signal level that is needed to trigger a detection of a pulse. *Maximum = 4095* *Minimum = 0* *Default = 0* Example: SYS:COMP:THR 100 .. image:: gnu.gif .. toctree:: :maxdepth: 2 :caption: Usage: